OPERATION MANUAL
Fraction collector LAMBDA OMNICOLL
Table of contents
1. SETTING UP THE FRACTION COLLECTOR
1.1 Assembling the fraction collector
1.2 Assembly of fraction collection tubing (single stream)
1.3 How to increase the capacity of the fraction collector?
1.4 Multi-channel fraction collection – multiple stream sampling
1.4.1 Standard multi-stream fraction collection
1.4.2 Multi-stream fraction collection VERSION 2022
1.5 Connecting the drop-counter detector (optional)
2. START OF THE FRACTION COLLECTION
3. PROGRAMMING OF THE FRACTION COLLECTOR
3.1 Principle of programming
3.2 Ready to use Fraction collector with the supplied tube racks
3.3 Programming the OMNICOLL for any tube racks or recipients
3.4 Control Panel of OMNICOLL Fraction Collector
3.5 Fraction collection according to the volume
3.6 Calibration of the peristaltic pump and the OMNICOLL
3.7 Fraction collection with a time interval in-between fractions (“high”)
4. REMOTE CONTROLS
4.1 Collection of single samples
4.2 Collection of multiple samples
8. APPENDIX
8.1 RS communication protocol for LAMBDA OMNICOLL fraction collector-sampler
8.2 How to set the OMNICOLL Fraction Collector-Sampler address?
8.3 RS-connection scheme
1. SETTING UP THE FRACTION COLLECTOR
1.1 Assembling the fraction collector
The setup of the fraction collection - sampler, LAMBDA OMNICOLL is very easy - a short video of the installation can be found at http://www.youtube.com/watch?v=33J9U_2-b-o
Figure 1.1-9 Place the control unit just at the beginning
of the gear rails. The distance from the edge should be
the same on the left and right sides. The gear wheels
of the control unit must engage correctly in the gears
of both rails. Be sure that nothing will disturb the movement
of the control unit during fractionation.
1.2 Assembly of fraction collection tubing
1.3 How to increase the capacity of the fraction collector?
Since several lower parts of the collector can be coupled together, the capacity of the collector can be increased many times.
The control unit can move freely from one lower unit to the other on the gear rails. Only two rectangular coupling bars are needed to keep two units together.
- Remove the two blind bars from the rear side of the frame of the first lower part unit and the front side of the second lower part unit. Insert the coupling bars (Art. No. 6912) on their place and fix with the four bolts.
- Be aware that you must remove the fixed stop signal from the first frame unit and place it onto the next one. You may also use the magnetic stop contact.
- Make sure that nothing will block the movement of the control unit during the fraction collection.
Figure 1.3-1 Fraction collector with extended capacity for 12 streams and 250 ml fraction collection bottles.
The tube capacity of the LAMBDA OMNICOLL fraction collector and sampler can be easily increased by the addition of lower support. Virtually any number of capacity extensions can be added.
1.4 Multi-channel fraction collection – multiple stream sampling
An interesting property of the LAMBDA OMNICOLL fraction collector is the possibility to collect simultaneously the streams of many chromatographic columns.
For LAMBDA OMNICOLL multi-stream fraction collectors, different designs exist:
1.4.1 Standard multi-stream fraction collection
The standard multi-stream assembly is available for 2 to 18 or even more effluents of multiple chromatography columns. This multi-stream assembly is technically very simple and polyvalent. It allows an easy adaptation of various experimental needs directly by the user.
Examples of the standard multi-stream fraction collection:
1.4.1.1 Mounting of the standard front multiple stream adapter
Figure 1.4.1-11 Place the tubing guide holder in position above the
first fraction of each stream (channel). If 4 streams are collected
in a row, the place: first tubing guide over tube no. 1, second
tubing guide over tube no. 4, third tubing guide over tube no. 7
and fourth tubing guide over tube no. 10
For more information about the multiple stream mounting, please do not hesitate to contact us support@lambda-instruments.com.
1.4.2 Multi-stream fraction collection VERSION 2022
LAMBDA OMNICOLL with accessories for multi-stream fraction collector VERSION 2022 can be carried out from 2 to 64 simultaneous threads. LAMBDA OMNICOLL accessories for multi-stream fraction collection VERSION 2022 are:- art. no. 6952 Accessory for fixed multi-stream collection adjustable up to 64 channels
- art. no. 6908-n Tube guide, 100 mm
- art. no. 6950 Accessory for mobile multi-stream collection adjustable up to 32 channels
- art. no. 6951 Accessory for mobile multi-row collection adjustable up to 8 channels
Figure 1.4.2‑1 Front side of the customized multi-channel fraction collector LAMBDA OMNICOLL for 18 streams. (Accessory for fixed multi-stream collection adjustable up to 64 channels art. no. 6952).
Figure 1.4.2-2 Back side of the customized multi-channel fraction collector LAMBDA OMNICOLL for 18 streams. (Accessory for fixed multi-stream collection adjustable up to 64 channels art. no. 6952).
Figure 1.4.2‑3 Configuration for collecting simultaneous streams into 96-well plates (art. no. 6951 - Accessory for mobile multi-row collection adjustable up to 8 channels)
Figure 1.4.2‑4 Customized multi-channel fraction collector LAMBDA OMNICOLL for 6 streams (art. no. 6950 - Accessory for mobile multi-stream collection adjustable up to 32 channels).
1.4.2.1 Mounting #6950: mobile multi-stream collection adjustable up to 32 channel
Figure 1.4.2‑5 First you need to assemble the moving arm holder.
Figure 1.4.2‑6 The holder is fixed with two screws using a hexagonal key.
Figure 1.4.2‑7 The moving arm holder attaches to the black rectangle attached to the device.
Figure 1.4.2‑8 The holder is fixed with one screw using a hexagonal key.
Figure 1.4.2‑9 The moving arm is mounted on the holder.
Figure 1.4.2‑10 The moving arm is fixed with one washer and bolt.
Figure 1.4.2‑11 The tube guide is mounted on a moving arm in accordance with the selected position.
(For each stream one tube guide (art. no. 6950) is needed).
Figure 1.4.2‑12 The tube guide is fixed with one bolt.
Figure 1.4.2‑13 The tube guide can be raised or lowered with a hexagonal key.
Figure 1.4.2-14 With LAMBDA OMNICOLL sold in 2023 and newer versions, the pipe guide can be raised or lowered with 1. bolt and mounted on the moving arm according to the selected position with 2. bolt without hexagonal keys.
1.5 Connecting the drop-counter detector (optional)
The drop counter is an optional device which could be obtained on request.
Fractionation can also be done according to the desired number of drops in the fraction collector using the drop-counter.
Figure 1.5-3 Partially unscrew the tubing guide holder screw (A),
so that the tubing guide (B) can be rotated. Slightly loosen the
nut (C), so that the PTFE tubing (D) can move freely inside the
tubing guide (B).Screw the drop counter (E) on the lower end of
the tubing guide (B).Tighten the tubing guide holder screw (A)
until the tubing guide (B) does not move freely. Securely tighten
the nut (C), to prevent the movement of PTFE tubing (D) inside
the tubing guide (B). Fix the drop counter cable (F) to the tubing
guide (B) with the help of Teflon or a spiral wrap, etc. Let the
drop-counter cable (F) pass together with the PTFE tubing (D)
through the magnetic holder (H).
2. START OF THE FRACTION COLLECTION
• Place the control unit just at the beginning of the gear rails. (Figure 1.1-9) The distance from the edge should be the same on the left and right sides. The gear wheels of the control unit must engage correctly in the gears of both rails. Be sure that nothing will disturb the movement of the control unit during fraction collection.
• Connect the cable of the control unit of the OMNICOLL fraction collector and sampler with the power supply plug. (Figure 1.1-10)
• Connect the power supply to the AC mains (95-240 V / 50-60 Hz). The control unit and its arm move automatically to the first position in the first row.
• Press the SET button (a short acoustic signal is heard). By pressing the COUNT/TIME button you can select fractionation according to time or volume. The yellow LED indicates which collection mode has been selected. (Refer section 3.4.1 also).
• Use the four buttons Λ Λ Λ Λ under the display to select the desired value. Confirm it by pressing the SET button again (a long acoustical signal is heard). The fraction time can be set from 0.1 min to 999.9 min (about 16.6 hours) in 0.1 min steps or from 1 to 9999 min in 1 min steps. To change the time resolution, please refer to section 3.4.2. For the setting of the volume, please refer to section 3.5.
• Pass the dispensing tubing (PTFE) through the tubing holder as describe in section 1.2 and tighten the tube holder in the collector arm so that the dispensing tubing is about 1 cm above the tubes or recipients.
• Place the tube rack on the fixing mat (Figure 1.1-8), so that the first tube is just under the tip of the tubing and the edge of the rack is parallel to the yellow frontal support (Figure 1.1-12). Place the additional tube racks immediately behind the first and following racks.
• Make sure that nothing will block the movement of the control unit during the fraction collection and that the tubing length is sufficient for taking all fractions or samples.
• Press the START button to begin the collection of fractions or sample taking (Figure 3.6-6). The green LED will be on and, if you use a LAMBDA peristaltic pump (PRECIFLOW, MULTIFLOW, HIFLOW or MAXIFLOW), it will be activated automatically. (refer section 3.6)
• You may use the magnetic block to stop the fraction collection. Place the supplied magnet on the right inner side of the frame (Figure 1.1-11)
It might happen that the position of the optical sensors lies outside the coding lines of the magnetic band inserted into the carriage.
When started in such a position, the fraction collecting arm will move either to the extreme left or right position, according to the instrument setting and is blocked there.
After several seconds an automatic protection will switch off the motor.
To move the arm into the coding mark signals just press the button STEP. The arm will move to the border coding line and will operate correctly within the marked limits of the inserted magnetic coded band.
If the moving arm is blocked in its movement or operated without inserted magnetic coding band, the motor will stop and the message “Eror” will appear on the display. Switch off the power, resolve the problem and start again as usual.
3. PROGRAMMING OF THE FRACTION COLLECTOR
A short video on programming the OMNICOLL fraction collector can be found at http://lambda-instruments.com/?pages=video-fraction-collector-and-sampler
3.1 Principle of programming
Laboratory practice shows that programming of instruments equipped with microprocessors are not simple and mistakes are easily made, especially when such instruments are used occasionally.
We have developed a new method, which should eliminate such problems and make the programming easy for all types of tube racks.
The white lines should be about 2 mm broad and full-length. These lines are detected by the photo-detectors of the OMNICOLL fraction collector and the fraction collector stops at these positions. The actual fraction position is about 1 mm after the first edge of the white lines.
The beginning and end signals (at Figure 3.1-3) determine at the same time the position of the first and the last fraction in the row. The fraction position is about 2 mm after the first edge of the white line of the beginning and end stripes.
This simple coding principle allows the use of any racks and fraction stands or other recipients with the LAMBDA OMNICOLL fraction collector and sampler.
3.2 Ready to use Fraction collector with the supplied tube racks
The magnetic coding bands programmed with white lines for the desired tube racks are provided with the OMNICOLL Fraction collector.
Place the magnetic coding strip strictly horizontally into the corresponding slot! Otherwise, the optical detectors could be damaged.
3.3 Programming the OMNICOLL for any tube racks or recipients
3.3.1 Coding X-axis for tube distance in row
Measure the distance between the tube centres of the desired rack and choose the number of fractions in a row.
The beginning and end signals at the lower or upper edge of the coding band informs the microprocessor whether the arm of the control unit is on right or left.
3.3.2 Coding Y-axis for row distance
Measure the distance between rows. When several tube racks are used, pay attention to difference in distance between the last row of the first rack and the first row of the second rack.
If the fraction collection should be stopped at the last row, then make the last line broader (approximately 1 cm thick). This functions as the stop signal.
Since the coding band is kept in place by magnetic force, it can be easily positioned.
The first row line of the Y-axis should be ca 10 cm from the front.
When the control unit is placed in the starting position on the front of the frame, the photoelectric detector on the left side of the control unit must be situated before the first line. Otherwise the control unit will stop at the position of the next row.
3.4 Control Panel of OMNICOLL Fraction Collector
3.4.1 Selection of linear, meander-like or row collection
The LAMBDA OMNICOLL fraction collector-sampler allows the following operation modes:
Line: | The fractions are collected from the left to the right. After the last fraction of the row, the collecting arm moves to the left to the first position of the next row. |
Meander: | The fractions are taken alternatively from the left to the right and in the following row from the right to the left and so forth. |
Row: | The collecting arm does not move within the row. The collector moves only from row to row. This collection mode is mainly used in the simultaneous (multi-channel) fraction collection mode. |
In the row mode, the position of the fraction collecting arm can be adjusted by using the STEP button.
A small difference in the fraction positions with respect to the tube centres may occur between fractions taken in opposite collection directions. If the fraction lines are not too broad (about 2 mm), this difference in position is of approx. 1 mm
3.4.2 Time resolution (0.1 or 1 minute)
The time resolution of the OMNICOLL fraction collector can be selected in the following way:
3.4.3 Drops and count divider (Division co-efficient)
This function allows increasing considerably (by 60 times) the volume of fractions either using a drop counter (art. no. 6926) or motor impulsions counter (counts).
The divider can be set either to divide by one or divide by 60. In the first setting, every drop and every impulsion of motor are counted. In the second setting, sixty drops or impulsions give just one signal.
The volume of fractions can be also varied by the selection of tubing of internal diameters from 0.5 to 4 mm, when constant volume fractions are collected by counting pump impulsions (in “COUN” operation mode – COUNT LED is on, see section 3.5).
3.4.4 Address selection (for PC control)
When the OMNICOLL fraction collector and sampler has been equipped with the optional RS-232 interface, it can be controlled digitally, e.g. from a PC.
3.5 Fraction collection according to the volume
If you use the OMNICOLL fraction collector together with LAMBDA peristaltic pumps PRECIFLOW, MULTIFLOW, HIFLOW or MAXIFLOW, you can take fractions of precise volume (from 0.05 to 500 ml or 0.6 to 30 litres per fraction).
The heart of these pumps is a stepping motor or BLDC motor, controlled by a generator of electric impulses (microprocessor). After each impulse the pump motor moves by one step. This movement displaces a very small and precise volume of liquid.
The collector counts these impulses and thus makes it possible to deliver an exact volume of liquid for each fraction. This method is an attractive alternative to the old drop counting procedure, where the volume is affected by the viscosity, surface tension etc. of the respective liquid.
Since the diameter of the tubing used in the peristaltic pumps affects the speed of the flow (flow rate), the pump has to be calibrated before use in order to establish the relation between the count number and the delivered liquid volume.
It is also possible to synchronize OMNICOLL fraction collector with third-party (e.g.: Cole parmer, Ismatec, Flexicon, etc.) multi-channel or single channel peristaltic pumps. We can offer customized remote control cable and communication module for pump switching and RS-232 connection (Art. No.: 6911) for the desired peristaltic pump connection.
3.6 Calibration of the peristaltic pump and the OMNICOLL
The LAMBDA peristaltic pump is connected to the OMNICOLL fraction collector via the remote control cable (Art. No. 4810-s) and the communication module box (Art. No. 6911 or Art. No. 6929). The plug-in power supply is also connected to this connection box. (It is impossible to make a wrong connection because the connectors will not fit to the wrong sockets).
Measure the volume of this fraction. This volume corresponds to the pre-set count number.
From this ratio (volume/count number) you can easily calculate the count number corresponding to the desired fraction volume.
Since tubing of 0.5 to 4 mm of internal diameter (with a tubing wall thickness ~1 mm) can be used in the LAMBDA peristaltic pumps, the fraction volume can be selected in a wide range. For example, when tubing with an internal diameter of 3 mm is used, then one count corresponds approximately to one drop.
3.7 Fraction collection with a time interval in-between fractions (“high”)
LAMBDA OMNICOLL fraction collector and sampler could be used for the collection of samples with a time interval between consecutive fractions varying between 0.1 min and 16.6 hours (999.9 min) or 1 min and 166 hours (9999 min).
This is particularly useful during fermentations and other biological and chemical processes. For this application, switch the microprocessor of the control unit to the “high” mode.
Selecting the high mode:
In the “high” mode you can successively select the parameters TIME, COUNT, PAUSE and NUMBER by pressing the COUNT/TIME button.
After selection of the parameter with the COUNT/TIME button, press the SET button.
After a short beep the name of the parameter will appear during one second, followed by last used value of the parameter.
You can change this value by means of the four buttons Λ Λ Λ Λ under the display. Press the button SET to confirm the new value (long beep).
After setting of the parameters, the fraction collection is started by pressing the START button.
The collector will take the number of samples as programmed under the NUMBER setting and will stop for the time programmed under the PAUSE setting. This cycle will go on until a STOP signal will be encountered. The LAMBDA peristaltic pump is activated automatically only during fraction collection.
When the fraction collection is controlled from the panel (i.e. not by remote control signal) the fraction collector will collect the number of fractions/samples introduced in the parameter NUMB of duration TIME or volume COUNT (if a drop counter or LAMBDA pumps are used), followed by the PAUSE, if a PAUSE was set.
Then, the fraction collector will repeat this till the STOP signal is encountered (i.e. STOP is pressed or the fraction collector is stopped by the end magnet). This collection mode allows for e.g. a line washing step before the relevant fraction is taken.
If operated by external (analog) remote control signal, the fraction collector will take the set number of fractions/samples of duration TIME or volume COUNT (if a drop counter or LAMBDA pumps are used), and will then stop till the reception of a new external remote control signal.
The parameter PAUSE is ignored in this operation mode.
Such a collection mode could be useful for taking of fractions/samples in a process where an alarm situation of a selected process parameter has occurred and the samples can be used for further analysis.
4. REMOTE CONTROLS
Sometimes it could be useful to take samples during a process, at a moment which is not known in advance.
For example, during fermentation (or any other biological, chemical or physical processes) certain limit values activates the alarm signal . This signal can be used to take a fraction. This fraction may be essential to recognize the reason for the alarm.
The fraction collector can take one or several samples after detection of a signal 3 to 12 V (or 12 to 30V with a 3300 ohm resistor switched in series). The LAMBDA peristaltic pump will be activated automatically.
For safety reasons the voltage of the external signal must not exceed 48 V to earth!
4.1 Collection of single samples
If the collector operates in normal modus, it will take a single fraction of selected time or volume after the reception of a signal impulsion. It will then move to the next position and wait the following signal. This operation will go on until the detection of a stop signal.
4.2 Collection of multiple samples
If the collector operates in the high modus, after detection of a signal it will take the number of fractions programmed under “NUMB”, e.g. from 1 to 999 (refer section 3.7).
This feature is particularly interesting, when it is necessary to wash tubing before taking the fresh significant sample. This is very important during fermentation processes or cell cultures, where the culture may be inactivated when kept in the tubing for a long period of time. The same is true for any other process where the sample quality might be affected by staying in the tubing
5. PRACTICAL ADVICE
• Keep all the magnetic coding strips with signal lines clean. If necessary, you can clean the old or damaged signal lines with mild soap and soft water and dry it. Then, draw new lines at the same position. Instead of using a white pen you may also use strips of white adhesive tape (~1.5 to 2 mm wide).
• Keep the fixing mat clean and dry. Humidity or dust will provide only low adhesion of the fixing mat. The fixing mat can be washed with water to regain the original adhesion power.
• Take only 10 fractions in a row, even though your stand allows more positions. It will help you to find the desired fraction more easily and you will eliminate mistakes.
• The alignment of the tube rack will be easier, if you press it against the front part of the frame supports. By pushing the magnetic Y – axis coding strip you will find the position corresponding to this rack position.
• By removing the tub sheet from the frame unit, it is possible to use even larger or taller collecting containers. We can also supply tub sheets to use with funnels or according to your needs. This makes it possible to use even largest containers for taking fractions. For such applications a modified tubing holder can also be made.
• On demand we can also supply an adapter for simultaneous collection of multiple fractions. This enables to collect simultaneously the streams of several chromatographic columns. (Refer section 1.4 for multi-stream operation instructions.)
6. MAINTENANCE
The LAMBDA OMNICOLL fraction collector does not require any special maintenance. Keep your fraction collector clean. If necessary, clean it with damp cloth. You may use neutral detergents or ethyl alcohol.
If you have any difficulties or questions concerning your OMNICOLL fraction collector, please contact our service office (support@lambda-instruments.com).
7. FOR YOUR SAFETY
Thanks to the use of a plug – in power supply giving only a low voltage of 9 V DC the danger of electrical shock during the use of the LAMBDA OMNICOLL fraction collector and sampler has been virtually eliminated.
If the fraction collector is not used for an extended period of time, disconnect it from the mains. A modern miniaturized switching power supply is used, which has only a negligible consumption of electric current when the fraction collector is not in use.
8. APPENDIX
8.1 RS communication protocol for LAMBDA OMNICOLL fraction collector-sampler
8.1.1 Format of data sent by the PC to the fraction collector and back
Data sent by the PC: | #ss mm a xxxx qs c |
Data sent back by the pump: | <mm ss a xxxx qs c |
where, | |
# | is the first sign of a command sent by PC |
< | is the first sign of a message sent by fraction collector |
ss | is the address of the fraction collector (slave) |
mm | is the address of the PC (master) |
A | is the command (see section 8.1.2) |
x | is the data value (4 ASCII numbers from 0 to 9; sent from the highest order digit to the lowest order digit) |
Qs | is the control sum in HEX format (2 ASCII characters of the type 0…9ABCDEF) |
C | is the end sign cr (carriage return) The collector will fulfil the task and block any manual command on the front panel. |
8.1.2 Commands
# ss mm r qs c | start (run) |
# ss mm e qs c | activates remote control of the collector (front panel deactivated) |
# ss mm g qs c | activates local mode (front panel activated) |
# ss mm s qs c | stopp |
# ss mm f qs c | step forward |
# ss mm b qs c | step back |
# ss mm w qs c | step in actual moving direction (depending on LINE or MEAN setting) [corresponds to pressing the STEP button] |
# ss mm l qs c | step to next line |
# ss mm h qs c | “high” mode |
# ss mm u qs c | “normal” mode |
# ss mm m qs c | “MEAN” collection mode (meander or zigzag collection mode) |
# ss mm v qs c | “LINE” collection mode (collects fractions always from left to right) |
# ss mm i qs c | “ROW” collection mode, the collector moves only from row to row |
# ss mm d qs c | unit setting – 0.1 minute step time setting (XXX.X) |
# ss mm j qs c | unit setting – minute step time setting (XXXX) |
# ss mm o qs c | open valve |
# ss mm c qs c | close valve |
# ss mm a qs c | division coefficient setting “1” |
# ss mm k qs c | division coefficient setting “1/60” |
# ss mm p xxxx qs c | number of pulses from pump or drop counter |
# ss mm t xxx.x qs c | collection time (in 0.1 minute steps) |
# ss mm t xxxx qs c | collection time (in minute steps) |
# ss mm q xxx.x qs c | pause time between two fractions (in 0.1 minute steps) (fraction collector automatically enters “high” mode) |
# ss mm q xxxx qs c | pause time between two fractions (in minute steps) (fraction collector automatically enters “high” mode) |
# ss mm n xxxx qs c | number of fractions (fraction collector automatically enters “high” mode) |
# ss mm G x qs c | to request the fraction collector to send data to the PC |
where x is a number from 0-3:
0: collection time setting (TIME)
1: pulse setting (COUNT)
2: pause time setting (PAUSE)
3: number of fractions setting (NUMBER)
The answer of the OMNICOLL fraction collector-sampler is the following:
< mm ss B xxx.x qs c | time xxx.x in 0.1 minute steps (collector is in stand-by mode) |
< mm ss B xxxx qs c | time xxxx in minute steps resp. number of pulses or fractions (collector is in stand-by mode) |
< mm ss R xxx.x qs c | time xxx.x in 0.1 minute steps (collector is in operation) |
< mm ss R xxxx qs c | time xxxx in minute steps resp. number of pulses or fractions (collector is in operation) |
8.1.3 Checksum control
The following examples show, how the checksum is calculated:
mm = 01 [PC address (master) is set to 01]
ss = 02 [Device address (slave) is set to 02]
The PC sends: #0201g4Dcr
The control addition (checksum) is made in the following way (only the last byte (2 ASCII characters of the type 0…9ABCDEF) is taken):
# | 0 | 2 | 0 | 1 | g | 4D (last byte) | cr |
23h | +30h | +32h | +30h | +31h | +67h | =14Dh | 0Dh |
The PC sends: #0201t102320cr
The control addition (checksum) is made in the following way (only the last byte (2 ASCII characters of the type 0…9ABCDEF) is taken):
#
23h |
0
+30h |
2
+32h |
0
+30h |
1
+31h |
t
+74h |
1
+31h |
0
+30h |
2
+32h |
3
+33h |
20 (last byte)
= 220h |
cr
0Dh |
8.1.4 Format of the data transmission
Speed: | 2400 Bd (Baud)8 data bits, odd parity, 1 stop bit |
The control addition (checksum) is made in the following way:
# | 0 | 2 | 0 | 1 | t | 1 | 0 | 2 | 3 | =2 | 2 | cr |
23h | +30h | +32h | +30h | +31h | +74h | +31h | 30h | +32h | +33h | =32h | 32h | 0dh |
8.2 How to set the OMNICOLL Fraction Collector-Sampler address?
When the LAMBDA OMNICOLL fraction collector and sampler has been equipped with the optional RS-232 interface, it can be controlled digitally, e.g. from a PC.
Disconnect the fraction collector-sampler from the mains. While keeping the direction button STEP pressed connect the fraction collector to the mains again. The message “A” and two numbers will appear on the display. This number from 00 to 99 is the current address of the fraction collector. To change the address press the buttons Λ Λ Λ Λ under the display until the desired number is obtained. Confirm your choice by pressing the button SET.
8.3 RS-connection scheme
The 5-pole DIN connector “REMOTE” of the communication module (Art. No. 6911 or 6929) is used for the remote control and RS connection.
When the optional RS-232 interface is available the pins are used as follows:
No. | Colour | Description |
1 | blue | Input remote control +3-12 V |
2 | white | RS-232 TTTL (TXD) |
3 | black | Ground (GND) |
4 | ||
5 | yellow | RS-232 RTTL (RXD) |